描述
Main Technical Parameters
(1) Power Supply: AC 220V ± 5% 50Hz
(2) Heating Power: 4000W
(3) Control Temperature: 80℃
(4) Working Environment Temperature of the Instrument: -10℃ to +45℃
(5) Environmental Temperature Rise: No more than 85%.
Usage Method
(1) Place an appropriate amount of water in the water tank, turn on the power supply, and set the control temperature of the instrument to 80℃. The instrument will stabilize at 80℃ after a period of time.
(2) Place the reaction tank on the shelf inside the refrigerator and level it.
(3) After 24 hours of constant temperature, remove the reaction tank, turn off the power supply.
(4) After the test is completed, open the water valve at the bottom of the box, drain the water inside the box, wipe off the moisture on the surface of the instrument, and place it in a ventilated and dry place.
Packing List:
Serial Number Name Quantity
1 Mainframe 1 unit
2 Instruction Manual 1 copy
3 Certificate of Conformity 1 copy
4 Micrometer Self-provided